Quadrupole ArtifactInfoSheet: - Artifacts - 
Case Studies, 
Reduction Index, 
etc.MRI Resource Directory:
 - Artifacts -
 
Quick Overview
Artifact Information
NAME
Quadrupole artifact
DESCRIPTION
Signal loss, intensity variations
REASON
B1 disturbance
HELP
Fat suppression (SPIR or FatSat) is very critical to the magnetic field homogeneity. Eddy currents in the patient results in B1 disturbance from left to right and from anterior to posterior. The artifact is seen as signal intensity variations with SPIR, like a signal intensity loss diagonal in the image. The short T1 inversion recovery (STIR) sequence is due to another type of fat suppression insensitive to this artifact.
mri safety guidance
Image Guidance
spacer
MRI Resources 
MRI Accidents - Guidance - Societies - Liver Imaging - Pathology - Non-English
 
Signa 3.0Tâ„¢InfoSheet: - Devices -
Intro, 
Types of Magnets, 
Overview, 
etc.MRI Resource Directory:
 - Devices -
 
www.gehealthcare.com/usen/mr/s_excite3/index.html (Signa VH/i 3.0T)
With GE Healthcare leading-edge technology in ultra-high-field imaging. The 3 T VH/i provides a platform for advanced applications in radiology, cardiology, psychology and psychiatry. Real-time image processing lets you acquire multislice whole brain images and map brain functions for research or surgical planning. And the 3 T Signa VH/i is flexible enough to provide clinicians with high performance they require. It can provide not only outstanding features in brain scanning and neuro-system research, but also a wide range of use in scanning breasts, extremities, the spine and the cardiovascular systems.
Device Information and Specification
CLINICAL APPLICATION
Whole body
CONFIGURATION
Cylindrical - high homogeneity
T/R quadrature head, T/R quadrature body, T/R phased array extremity (opt)
SYNCHRONIZATION
ECG/peripheral, respiratory gating
PULSE SEQUENCES
SE, IR, 2D/3D GRE, FGRE, RF-spoiled GRE, FSE, Angiography: 2D/3D TOF, 2D/3D phase contrast vascular
IMAGING MODES
Single, multislice, volume study, fast scan, multi slab, cine, localizer
SINGLE SLICE
100 Images/sec with Reflex100
MULTISLICE
100 Images/sec with Reflex100
1 cm to 40 cm continuous
2D 0.5-100mm in 0.1mm incremental
1280 x 1024
MEASURING MATRIX
128x512 steps 32 phase encode
PIXEL INTENSITY
256 gray levels
55cm
MAGNET WEIGHT
15102 kg incl. cryogen's
H*W*D
260cm x 238cm x 265cm
POWER REQUIREMENTS
480 or 380/415, 3 phase ||
COOLING SYSTEM TYPE
Closed-loop water-cooled grad.
Less than 0.14 L/hr liquid He
STRENGTH
40mT/m
5-GAUSS FRINGE FIELD, radial/axial
5.4 m x 3.2 m
Superconductive + hi order active
spacer

• View the DATABASE results for 'Signa 3.0T™' (2).Open this link in a new window

MRI Resources 
Directories - Lung Imaging - Cardiovascular Imaging - Brain MRI - Fluorescence - Nerve Stimulator
 
Gradient EchoForum -
related threads
 
(GE) An echo signal generated from a free induction decay by means of a bipolar switched magnetic gradient. The echo is produced by reversing the direction of a magnetic field gradient or by applying balanced pulses of magnetic field gradient before and after a refocusing RF pulse so as to cancel out the position dependent phase shifts that have accumulated due to the gradient.
In the latter case, the gradient echo is generally adjusted to be coincident with the RF spin echo. When the RF and gradient echoes are not coincident, the time of the gradient echo is denoted echo time (TE) and the difference in time between the echoes is denoted time difference (TD).
Gradient echo does not refocus the effects of main field inhomogeneity and therefore is generally used with a short echo time. Disadvantages of gradient echo imaging are compromised anatomic details and artifacts in regions with varying susceptibility e.g. between the air-containing sinuses and brain and especially between haemorrhages and normal tissue.

See also Susceptibility Artifact.
spacer

• View the DATABASE results for 'Gradient Echo' (121).Open this link in a new window

 
Further Reading:
  Basics:
Mapping of low flip angles in magnetic resonance(.pdf)
Saturday, 1 January 2011   by www.hal.inserm.fr    
A LANTHANIDE LANTHOLOGY(.pdf)
   by www.phy.davidson.edu    
Enhanced Fast GRadient Echo 3-Dimensional (efgre3D) or THRIVE
   by www.mri.tju.edu    
Searchterm 'Field Homogeneity' was also found in the following service: 
spacer
Forum  (1)  
 
Magnetic Resonance SpectroscopyMRI Resource Directory:
 - Spectroscopy pool -
 
(MRS / MRSI - Magnetic Resonance Spectroscopic Imaging) A method using the NMR phenomenon to identify the chemical state of various elements without destroying the sample. MRS therefore provides information about the chemical composition of the tissues and the changes in chemical composition, which may occur with disease processes.
Although MRS is primarily employed as a research tool and has yet to achieve widespread acceptance in routine clinical practice, there is a growing realization that a noninvasive technique, which monitors disease biochemistry can provide important new information for the clinician.
The underlying principle of MRS is that atomic nuclei are surrounded by a cloud of electrons, which very slightly shield the nucleus from any external magnetic field. As the structure of the electron cloud is specific to an individual molecule or compound, then the magnitude of this screening effect is also a characteristic of the chemical environment of individual nuclei.
In view of the fact that the resonant frequency is proportional to the magnetic field that it experiences, it follows that the resonant frequency will be determined not only by the external applied field, but also by the small field shift generated by the electron cloud. This shift in frequency is called the chemical shift (see also Chemical Shift). It should be noted that chemical shift is a very small effect, usually expressed in ppm of the main frequency. In order to resolve the different chemical species, it is therefore necessary to achieve very high levels of homogeneity of the main magnetic field B0. Spectra from humans usually require shimming the magnet to approximately one part in 100. High resolution spectra of liquid samples demand a homogeneity of about one part in 1000.
In addition to the effects of factors such as relaxation times that can affect the NMR signal, as seen in magnetic resonance imaging, effects such as J-modulation or the transfer of magnetization after selective excitation of particular spectral lines can affect the relative strengths of spectral lines.
In the context of human MRS, two nuclei are of particular interest - H-1 and P-31. (PMRS - Proton Magnetic Resonance Spectroscopy) PMRS is mainly employed in studies of the brain where prominent peaks arise from NAA, choline containing compounds, creatine and creatine phosphate, myo-inositol and, if present, lactate; phosphorus 31 MR spectroscopy detects compounds involved in energy metabolism (creatine phosphate, adenosine triphosphate and inorganic phosphate) and certain compounds related to membrane synthesis and degradation. The frequencies of certain lines may also be affected by factors such as the local pH. It is also possible to determine intracellular pH because the inorganic phosphate peak position is pH sensitive.
If the field is uniform over the volume of the sample, "similar" nuclei will contribute a particular frequency component to the detected response signal irrespective of their individual positions in the sample. Since nuclei of different elements resonate at different frequencies, each element in the sample contributes a different frequency component. A chemical analysis can then be conducted by analyzing the MR response signal into its frequency components.

See also Spectroscopy.
spacer

• View the DATABASE results for 'Magnetic Resonance Spectroscopy' (8).Open this link in a new window


• View the NEWS results for 'Magnetic Resonance Spectroscopy' (3).Open this link in a new window.
 
Further Reading:
  News & More:
Accuracy of Proton Magnetic Resonance Spectroscopy in Distinguishing Neoplastic From Non-neoplastic Brain Lesions
Saturday, 2 December 2023   by www.cureus.com    
MRI Resources 
MRI Reimbursement - MRI Technician and Technologist Schools - Cochlear Implant - Distributors - Case Studies - Corporations
 
Periodically Rotated Overlapping Parallel Lines with Enhanced ReconstructionInfoSheet: - Sequences - 
Intro, 
Overview, 
Types of, 
etc.
 
(PROPELLER) The PROPELLER MRI technique reduces the sensitivity to various sources of image artifacts (e.g., motion artifact, field inhomogeneity artifact, eddy current artifact). PROPELLER can be used with gradient echo and turbo spin echo sequences in a wide range of applications to improve the image quality, for example cardiac MRI, brain MRI, and pediatric examinations.
spacer
 
Further Reading:
  Basics:
Periodically Rotated Overlapping ParallEL Lines with Enhanced Reconstruction(PROPELLER) MRI; Application to Motion Correction
1999   by cds.ismrm.org    
MR Field Notes
   by www.gehealthcare.com    
Advances in Magnetic Resonance Neuroimaging
Friday, 27 February 2009   by www.ncbi.nlm.nih.gov    
  News & More:
Patient movement during MRI: Additional points to ponder
Tuesday, 5 January 2016   by www.healthimaging.com    
New MR sequence helps radiologists more accurately evaluate abnormalities of the uterus and ovaries
Thursday, 23 April 2009   by www.eurekalert.org    
MRI Resources 
Sequences - Contrast Enhanced MRI - Developers - Collections - Functional MRI - Lung Imaging
 
previous      11 - 15 (of 24)     next
Result Pages : [1]  [2 3]  [4 5]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



MRI is trending to low field magnets :
reduced costs will lead to this change 
AI will close the gap to high field 
only in remote areas 
is only temporary 
never 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 2 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]